1. Apply the fibrinogen solution to the tissue site which requires sealing and overlay with the thrombin solution.
2. Hold the tissues requiring sealing in place for a few minutes until a fibrin clot begins to form and adhesion is achieved.
For joint application of the fibrinogen and thrombin solutions, the application kit supplied in the Beriplast® P Combi-Set package can be used:
1. Insert the blue Y-piece into the white syringe holder. The fixation device should be bent at 90 in the middle. Insert the filled application syringes (remove needles) firmly into the Y-piece. Snap both syringes into the syringe holder. Attach the grip plate to the syringe plungers to prevent jamming of the plungers and to ensure smooth forward movement.

2. It is easier to assemble the Pantaject® system if, after the Y-piece is attached to the syringe holder, the device is bent at 90 in the middle.
3. The syringes should then be pushed securely into the Y-piece before the syringe holder is snapped into position.
4. Fix a suitable device (eg needle) into the Luer-Lok® fitting. To ensure that the system is free from blockages, hold the Pantaject® so that the needle is pointing upward.
5. Beriplast® P Combi-Set is now ready for use. Apply the fibrin sealant to the site by gently pushing down on the plunger connector. NB: to prevent the needle from becoming blocked, do not attach it to the Pantaject® until immediately before use.
1. Assemble the Pantaject® and check that the fibrinogen and thrombin solutions have not become mixed at the head of the Y-piece causing a blockage. Keeping the Pantaject® upright, attach the spray-tip to the Luer-Lok® fitting of the Y-piece. Apply Beriplast® P Combi-Set by pushing down slowly and evenly on the plunger connector to produce a steady vapor of fibrin sealant. To prevent the spray-tip from becoming blocked, do not attach to the Pantaject® until immediately prior to use. If the spray-tip becomes blocked, replace it with a new spray-tip.
2. Holding the spray-tip 10-20 cm from the surgical site push down slowly and evenly on the plunger connector to produce a fine, even spray of Beriplast® P Combi-Set. A fine film of fibrin sealant will form on the wound site.

1. Insert the spray-head (1) in the conical recess of the syringe holder.
2. Connect the syringes (2) filled with the fibrinogen solution (blue) and thrombin solution (red) with the spray-head. Snap both syringes into the syringe holder (3). Connect the grip plate (4) to the syringe plungers to prevent jamming of the plungers.
3. Attach the gas tube to the spray-head with the connector (5) and ensure that the clip (6) is closed.

1. Remove the needles from the syringes filled with the fibrinogen solution (blue marking, 3) and thrombin solution (red marking, 4).
2. Insert the Y-piece (1) of the Catheject™ application device in the conical recess of the syringe holder (2), which should be bended at an angle of approx. 45 degrees.
3. Firmly connect both syringes to the Y-piece. The syringe with the fibrinogen solution (blue marking, 3) should be connected to the opening on the Y-piece marked with the F-indicator.
4. Snap both syringes into the syringe holder (2).
5. Connect the grip plate (5) to the syringe plungers by holding the Pantaject® at the syringe plungers. The grip plate prevents jamming of the syringe plungers and ensures smooth forward movement.